EVS has historically been told the same thing about the anesthesia cart: clean the sides, the back and the wheels, and nothing else. Those surfaces on the machine cart have no controls and no settings, which is why EVS can clean them. Everything else belongs to the anesthesia tech. I want to explain why this rule has existed, because the reason is a good one and it protects the patient.
What the anesthesia cart is
The anesthesia cart is the machine itself, with the monitor and the breathing circuit. In many ORs a separate supply cart sits beside it, holding airway and IV supplies and, depending on the hospital, medications, but this post is about the machine. The FDA regulates an anesthesia gas machine as a Class II medical device, because it is used to administer a general inhalation anesthetic to a patient and to maintain the patient’s ventilation (U.S. Food and Drug Administration, 2026). It has flow controls for oxygen, air and nitrous oxide, a vaporizer with a dial that sets how much anesthetic the patient receives, an oxygen flush valve, an adjustable pressure-limiting valve, a ventilator with its settings, and alarms (King and Vo, 2026).
Why cleaning it matters
The anesthesia work area is one of the places where contamination moves during a case. The SHEA expert guidance on infection prevention in the anesthesia work area (Munoz-Price et al., 2019) recommends that the facility clean and disinfect the high-touch surfaces on the anesthesia machine and work area between OR uses, with a hospital disinfectant that is compatible with the equipment, based on the manufacturers’ instructions for use. The same guidance lists the gas flow controls, the vaporizer dials and the adjustable pressure-limiting valve among the components particularly likely to be contaminated. Loftus et al. (2008) sampled 61 operative suites and found that bacterial contamination of the anesthesia work area increased over the course of a case, and that bacteria, including vancomycin-resistant enterococcus, reached previously sterile IV stopcock sets in 32 percent of cases. SHEA also notes that few studies have evaluated specific cleaning products or practices for this area. The surfaces involved are the ones the provider touches all day, and the cleaning is not optional. The question is who does it.
Why EVS does not do it
Cleaning those surfaces means handling the controls. If we wipe the flow knobs and the vaporizer dial, we would absolutely have to turn some of them, and a control that is moved and not returned to the right setting changes what the patient receives. There are a lot of controls, and the anesthesia tech who comes in next would have to check every one of them to catch it. A patient could be harmed, and in the worst case could die. That is the reason, and I do not need a case to make the point. I will also say that I have not found a published case in which a cleaner moved a control and a patient was harmed. The argument rests on how the device works and on what the safety data says about gas delivery equipment.
The American Society of Anesthesiologists Closed Claims Project review of gas delivery equipment (Mehta et al., 2013) found that 85 percent of the 40 claims from 1990 to 2011 involved provider error, and that 35 percent were judged preventable by a preanesthesia machine check. A check helps, and it is not a guarantee.
Manufacturers publish cleaning instructions in their operator manuals, and the SHEA guidance tells facilities to disinfect based on the manufacturers’ instructions for use. What the manuals I found give is the how and not the who. Dräger’s Apollo manual (Dräger, n.d.), for example, says only products from its list of surface disinfectants may be used. Approved products differ by machine, and a standard EVS wipe may not be on the list for a given model, which is a second reason the work belongs with people trained on the specific model.
The fair argument against my position
The strongest argument against me is a fair one. EVS are the trained cleaning and disinfection professionals, and nurses and techs are often assumed to know how to disinfect without ever being taught. A study in Anesthesia & Analgesia by Goebel et al. (2016) monitored 200 cleaning and disinfection procedures on anesthesia workplaces, done either by anesthesia nurses or by specially trained housekeeping staff, and found that the housekeeping staff were faster (1.2 minutes against 2.6) and achieved a larger decrease in bacterial load than the anesthesia nurses. I accept that result. It tells me training is what makes cleaning work, and I agree with that. What it does not test is who should handle a live device with controls. Who cleans best and who may touch a regulated machine are two separate questions, and the study answers only the first.
If a hospital wanted EVS to take on more of the machine, the conditions are plain: training on each machine model, the manufacturer-approved product for that machine, a clear rule about which controls are touched and which are not, and someone accountable for confirming the setup afterward. Until those exist, the sides, the back and the wheels of the machine cart are the right job for EVS, because they have no controls or settings to disturb. The criticism runs the other direction too. If anesthesia techs are assumed to know how to disinfect, they should be trained and audited the same way EVS is.
I believe the person who cleans a device with live controls has to be trained on those controls and accountable for leaving the device as it was found. That is the anesthesia tech. EVS cleans what can be cleaned without changing what the patient receives, and the anesthesia tech cleans the rest.
References
Dräger (n.d.) Apollo operating instructions, software 3.2n, chapter 11: Cleaning and maintenance. Lübeck: Dräger. Available at: https://www.manualslib.com/manual/1807261/Dr-Ger-Apollo.html (Accessed: 22 November 2026).
Goebel, U. et al. (2016) ‘Bacterial contamination of the anesthesia workplace and efficiency of routine cleaning procedures: a prospective cohort study’, Anesthesia & Analgesia, 122(5), pp. 1444-1447. Available at: https://doi.org/10.1213/ANE.0000000000001220.
King, S. and Vo, C. (2026) Anesthesia machines: an overview. OpenAnesthesia. Available at: https://www.openanesthesia.org/keywords/anesthesia-machines-an-overview/ (Accessed: 18 November 2026).
Loftus, R.W. et al. (2008) ‘Transmission of pathogenic bacterial organisms in the anesthesia work area’, Anesthesiology, 109(3), pp. 399-407. Available at: https://doi.org/10.1097/ALN.0b013e318182c855.
Mehta, S.P. et al. (2013) ‘Patient injuries from anesthesia gas delivery equipment: a closed claims update’, Anesthesiology, 119(4), pp. 788-795. Available at: https://doi.org/10.1097/ALN.0b013e3182a10b5e.
Munoz-Price, L.S. et al. (2019) ‘Infection prevention in the operating room anesthesia work area’, Infection Control & Hospital Epidemiology, 40(1), pp. 1-17. Available at: https://doi.org/10.1017/ice.2018.303.
U.S. Food and Drug Administration (2026) 21 CFR 868.5160: Gas machine for anesthesia or analgesia. Electronic Code of Federal Regulations. Available at: https://www.ecfr.gov/current/title-21/section-868.5160 (Accessed: 18 November 2026).